4 Connect the medical device
4.1 Connecting the medical device to the water system
Ask an approved dental installation technician to connect your medical device to the water system.
The water supply system pressure may vary throughout the day. The water supply system pressure must be adapted
to the values recommended for your medical device. It is very important to make sure that the maximum pressure
permitted for the medical device is never reached or exceeded. If in doubt, you are strongly advised to install or
arrange for installation of a water pressure limiting system.
Before a long period of absence or when the medical device is not in use for a long period of time, turn off the water
supply to the device or dental practice.
4.2 Connecting the medical device to the water system
The supply pipe connector is used to connect the medical device to the domestic water distribution system. The
connector is extended by a pipe to which you must attach a filter.The filter needs to be cleaned and replaced regularly
as specified in the chapter Clean the water filter cartridge. page 21.
The water supply must comply with the quality criteria compatible with the practice of dental treatments.
4.3 Installing a water filter
Water supplied by the system in place may not be of a quality suitable for dental treatments.In this case, you must
install a water filter on the medical device's water supply pipe.
Procure the relevant kit and the installation and maintenance instructions for water and air filters and cartridges
[J11330].
SATELEC, a company of Acteon group accepts no responsibility for damage caused by an installation provided
by a service provider that has not been approved by us.
4.4 Connecting the medical device to the electrical network
Have your medical device connected to the mains power by an approved dental installation technician.
Check that the mains voltage is compatible with that indicated on the medical device or its mains adapter.
A different voltage would cause damage to the medical device and could injure the patient and the user. Any variation
in the electrical network voltage or electromagnetic field that is non-compliant with the limits in force, could interfere
with the medical device's operation.
Medical devices equipped with a protective earth must be connected to a supply network equipped with a protective
earth.
Do not plug the medical device into an extension lead and do not put the mains cord in a cable cover or cable
tray.