EasyManua.ls Logo

Acteon Newtron P5 - Monitoring and Routine Maintenance; Irrigation System Cleaning; Water Filter Cartridge Maintenance

Acteon Newtron P5
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Monitoring and routine maintenance
The only preventive maintenance the medical device requires is:
l Monitoring of accessories
l Routine cleaning, disinfection and sterilisation
l Cleaning
The medical device requires cleaning of the water filter cartridge.
Check the cleanliness of the air inlets on the control unit to prevent any heating.
Check the cleanliness of the handpiece nosepiece. It must be clean, smooth and corrosion-free. The handpiece must
screw easily and firmly inside it.
Check the condition of the handpiece rear seals, which must not be distended, torn or broken.
Before and after use, check the medical device and its accessories entirely for any problems. This is necessary to
detect any electrical isolation fault or damage. If necessary, replace damaged parts.
9.1 Cleaning the irrigation system
Operate the device at minimum power, at maximum irrigation flow rate for two minutes.
When the irrigation system has been cleaned, perform the following operations:
1. Disconnect the handpiece and refer to the handpiece cleaning, disinfection and sterilisation instructions
[J12911].
2. Clean and disinfect the medical device as indicated in the chapter Clean and disinfect the medical device page 19.
3. Follow the instructions for cleaning, disinfecting and sterilising SATELEC, a company of Acteon group
accessories [J81001] and [J02001].
9.2 Clean the water filter cartridge.
The water filter cartridge must be cleaned regularly. The water filter must be replaced every six months.
To clean the water filter cartridge, follow the steps below:
1. Stop the medical device (position O).
2. Unplug the electrical supply.
3. Shut off the water supply.
4. Disconnect the pipe from the device.
5. Unscrew the two filter sections.
User Manual | Newtron
®
P5 | J61101 | V6 | (13)|09/2017 | NBACEN030F - Page 21/38

Table of Contents

Related product manuals