Possible causes Solutions
Faulty handpiece seal
Replace the handpiece seal using the purpose-provided
kit.
10.4 Ultrasounds not working
Symptoms: the tip does not vibrate.
Possible causes Solutions
Tip loose
Fasten the tip using the wrench
Replace your torque wrench (once a year)
Faulty connector contact Clean the cord contacts
Handpiece cord wire(s) cut
Return to the Acteon Customer Service team to replace
the cord
10.5 Water leakage
Symptoms: water is leaking between the base of the handpiece and its cord.
Possible causes Solutions
Wear of 1.15 mm x 1 mm handpiece seal Replace the seal using the purpose-designed kit.
Page 26/38 - User Manual | Newtron
®
P5 | J61101 | V6 | (13)|09/2017 | NBACEN030F