EasyManuals Logo

Advanced Instruments Anoxomat Mark II CTS User Manual

Advanced Instruments Anoxomat Mark II CTS
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
35
Anoxomat Mark II CTS User’s guide
Appendix A Troubleshooting table
Error Cause Solution
Software freeze and customer is
not sure whether the jar has
been processed
Software error
Restart the system, open jar, vent and recycle the jar to
make sure desired environment has been created.
If the wrong tank is connected User error
When gas tank is replaced, incubate a quality control
plate before processing any patient sample to avoid any
loss of patient sample.
Printer does not function
Power connection is loose Tighten connection
Serial cable is loose Re-seat connection
No communication with
external computer
Cable is unplugged Plug in cable
Ethernet cable is damaged Replace cable
Ethernet connector is damaged Service the instrument to replace connector
Barcode scanner does
not function
Cable is unplugged
Secure cable. Note: If scanner does not function, you can
manually enter IDs and other relevant information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Anoxomat Mark II CTS and is the answer not in the manual?

Advanced Instruments Anoxomat Mark II CTS Specifications

General IconGeneral
BrandAdvanced Instruments
ModelAnoxomat Mark II CTS
CategoryMedical Equipment
LanguageEnglish

Related product manuals