Product specifications
Electrical
Instrument type 110V 220V
Voltage 110-120VAC 220-240 VAC
Frequency 50/60 Hz 50/60 Hz
Fuses (2) 6.3 A Slow 6.3 A Slow
Power consumption 516 W 516 W
Memory backup 2 Integral Lithim cells; 10-year life
(not user-serviceable)
Gas consumption (standard conditions in a jar suitable
for 12 petri dishes
Standard recipe
Microaerophilic Approximately 2 Liter (0.07 cubic
feet) per jar
Anaerobic Approximately 7 Liter
(0.25 cubic feet) per jar
Jars per cylinder Microaerophilic Anaerobic
10 Liter/150 bar >700 >200
10 Liter/200 bar >200 >250
50 Liter/ 150 bar >3600 >1000
50 Liter/200 bar >4900 >1400
Operating conditions
Temperature 10 °C to 32 °C, 50 °F to 91 °F
Humidity 20-80% relative humidity
(non-condensing)
Dimensions inches centimeters
Width 12.8 32.0
Depth 20.8 52.0
Height 11.2 28.0
Weight Pounds Kilograms
Net: 33.0 15.0
Shipping: 40.0 18.1
Warranty
One-year limited warranty on workmanship and
all parts except for glass, plastic, and parts warranted
by their makers.
Certification
Installation class: I
Over-voltage category: II
Pollution degree: 2
Moisture protection: IPX0 (ordinary)