273 ADAM-4000 Series User Manual
Chapter 8 Calibration
3. Set the resistance to “0” and click on “Save”.
4. Click on “Zero Cal.”
5. Apply 800.0 ohms resistor with 0.01% accuracy to CH0 and then click on “Save”