38
Aerogen
®
Table 11. Guidance and manufacturer’s declaration - electromagnetic immunity for
the Aerogen Pro nebulizer system that is not life supporting
This Aerogen Pro nebulizer system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Pro nebulizer
system should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment - Guidance
Conducted RF
IEC 61000-4-6
EN 61000-4-6:
2014
3 Vrms outside
industrial,
scientific and
medical (ISM) and
amateur radio
bands. 6 Vrms in
ISM and amateur
radio bands
150 kHz to 80 MHz
10 Vrms
150 kHz to 80
MHz
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Aerogen Pro nebulizer
system, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d = [1.17] √P