Aerogen
®
Pro System Instruction Manual
39
Table 11. Guidance and manufacturer’s declaration - electromagnetic immunity for
the Aerogen Pro nebulizer system that is not life supporting (Continued)
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment - Guidance
Radiated RF
IEC 61000-4-3
EN 61000-4-3:
2010
10 V/m
80 MH z to 2.7 G H z
27 V/m, 18 Hz PM
385 MHz
28 V/m, 50 %18
Hz PM
450 MHz
9 V/m, 217 Hz PM
710 MHz
9 V/m, 217 Hz PM
745 MHz
9 V/m, 217 Hz PM
780 MHz
28V/m, 18 Hz PM
810 MHz
28 V/m, 18 Hz PM
870 MHz
28 V/m, 18 Hz PM
930 MHz
28V/m, 217 Hz PM
1720 MHz
28 V/m, 217 Hz PM
1845 MHz
28 V/m, 217 Hz PM
1970 MHz
27 V/m, 217 Hz PM
2450 MHz
9V/m, 217 Hz PM
5240 MHz
10 V/m
80 MHz to
2.7 GHz
27 V/m,
18 Hz PM
385 MHz
28 V/m, 50 %18
Hz PM
450 MHz
9 V/m,
217 Hz PM
710 MHz
9 V/m,
217 Hz PM
745 MHz
9 V/m,
217 Hz PM
780 MHz
28V/m,
18 Hz PM
810 MHz
28 V/m,
18 Hz PM
870 MHz
28 V/m,
18 Hz PM
930 MHz
28V/m,
217 Hz PM
1720 MHz
28 V/m,
217 Hz PM
1845 MHz
d = [1.17] √P... 80MHz to 800MHz
d = [2.33] √P... 800MHz to 2.5GHz
where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,ª should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol: