10
Aerogen
®
Only use the Aerogen nebulizer technology with components specified
in the instruction manuals. Use of the Aerogen nebulizer technology with
components other than those specified in the Instruction Manual may
result in increased emissions or decreased immunity of the nebulizer
system.
Do not place the Aerogen USB Controller in an incubator during use.
To avoid exhaled medication affecting the ventilator, follow ventilator
manufacturer’s recommendations for use of a bacterial filter in the
expiratory limb of a breathing circuit.
Do not use in the presence of flammable substances or flammable
anesthetic mixtures combined with air, oxygen or nitrous oxide.
To avoid the risk of fire do not use to aerosolize alcohol-based medications,
which can ignite in oxygen-enriched air under high pressure.
Do not modify this equipment without the authorization of the manufacturer.
Inspect all parts before use, and do not use if any parts are missing,
cracked or damaged. In case of missing parts, malfunction or damage,
contact your sales representative.
Do not immerse or autoclave the Aerogen USB Controller or Aerogen USB
Controller AC/DC Adapter.
Do not microwave any parts.
Do not use or store outside of specified environmental conditions.
Follow local laws and recycling plans regarding disposal or recycling of
components and packaging.
Do not use in the presence of devices generating high electromagnetic
fields such as magnetic resonance imaging (MRI) equipment.