EasyManua.ls Logo

Aerogen USB Controller - Functional Test for Home Use

Aerogen USB Controller
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Aerogen
®
USB Controller System Instruction Manual
53
Cleaning of the Aerogen USB
Controller System
Table 3. Cleaning Summary
Product Wipe Clean Cleaning
Aerogen USB
Controller
See instructions below.
Aerogen Solo
The Aerogen Solo and
accessories are single
patient use only and
are not intended to be
cleaned, disinfected or
sterilized.
Aerogen Pro
See instructions below.
Aerogen USB Controller, controller cables and Aerogen USB Controller
AC/DC Adapter.
1. Wipe clean with an alcohol based disinfectant wipe or a quaternary
ammonium compound based disinfectant wipe.
2. Check for exposed wiring, damaged connectors, or other defects and
replace if any are visible.
3. Visually inspect for damage and replace the Aerogen USB Controller if
any damage is observed.
Note: Do not spray liquid directly onto the Aerogen USB Controller.
Do not immerse the Aerogen USB Controller in liquid.

Related product manuals