46
Aerogen
®
System Warnings
Read and study all instructions before using the Aerogen USB
Controller System.
If this product is being used to treat a life threatening condition, a backup
device is necessary.
During use observe for correct functioning of the nebulizer by regularly
verifying aerosol is visible and that no amber indicator lights are illuminated.
Do not use a filter or heat-moisture exchanger (HME) between the nebulizer
and patient airway.
The Aerogen nebulizers, T-pieces and accessories are not sterile.
The components and accessories of the Aerogen USB Controller System
are not made with natural rubber latex.
Only use physician-prescribed solutions that are approved for use with
a general purpose nebulizer. Consult drug manufacturer’s instructions
regarding suitability for nebulization.
Only use the Aerogen nebulizer technology with components specified
in the instruction manuals. Use of the Aerogen nebulizer technology with
components other than those specified in the Instruction Manual may
result in increased emissions or decreased immunity of the nebulizer
system.
To avoid exhaled medication affecting the ventilator, follow ventilator
manufacturer’s recommendations for use of a bacterial filter in the
expiratory limb of a breathing circuit.
Do not use in the presence of flammable substances or flammable
anesthetic mixtures combined with air, oxygen or nitrous oxide.