25
Cleaning of the Aerogen USB
Controller System
Table 3. Cleaning Summary
Product
Wipe
Clean
Disinfect
Sterilize
(Autoclave)
Aerogen USB
Controller
See instructions
below.
Aerogen Solo
The Aerogen Solo and
accessories are single
patient use only and
are not intended to be
cleaned, disinfected
or sterilized.
Aerogen Pro
See instructions
below.
Aerogen USB Controller, controller cables and Aerogen USB Controller
AC/DC Adapter.
1. Wipe clean with an alcohol based disinfectant wipe or a quaternary
ammonium compound based disinfectant wipe.
2. Check for exposed wiring, damaged connectors, or other defects and
replace if any are visible.
3. Visually inspect for damage and replace the Aerogen USB Controller if
any damage is observed.
Note: Do not spray liquid directly onto the Aerogen USB Controller.
Do not immerse the Aerogen USB Controller in liquid.