20
Aerogen
®
Continuous Nebulization Tube Set
The Aerogen Continuous Nebulization Tube Set is an accessory specific
to the Aerogen Solo nebulizer which enables safe continuous infusion of
liquid medication for aerosolization.
Note: Place the syringe cap on the syringe after it is filled with medication.
Figure 5. Continuous Nebulization Tube Set
1. Ensure the Aerogen Solo nebulizer is firmly fitted into the Aerogen Solo
T-piece in the breathing circuit.
2. Remove the syringe cap from the medication-filled syringe.
3. Attach the syringe end of the tubing onto the syringe.
4. Prime the tubing until the medication reaches end of tubing (Point A).
Note: The tubing priming volume is maximum 3.65 mL.
5. Unplug the tethered silicone plug from the Aerogen Solo nebulizer, but
do not remove it from the nebulizer.
6. Screw the nebulizer end of the tubing onto the top of the nebulizer.
A
Syringe Cap
Syringe
Tubing
(Syringe End)
Tubing
Tethered Silicone Plug
Tubing
(Nebulizer End)