49
Figure 2. Aerogen Palladium Vibrating Mesh Technology (Aerogen Vibronic
®
)
Use of the Aerogen Solo and T-piece during the administration of volatile
anesthetics may result in adverse effects on the constituent plastics. Do
not use with volatile anesthetics unless known to be compatible. Aerogen
have determined that, using anesthetic ventilators, the following volatile
anesthetic agents are compatible under the stated conditions below:
Anaesthetic Agent Proprietary Name
Maximum
Percentage of
Anaesthetic
Maximum Duration
of Exposure
Isoflurane FORANE
®
3.5 % 12 hours
Sevoflurane SEVOFLURANE
®
8 % 12 hours
Desflurane SUPRANE
®
10 % 12 hours