12
7) Move the scan up and down the entire length of the plumb line. If the
beam is slanted, and not vertical like the plumb line, the Z axis needs to
be calibrated.
Calibrating Z Axis
The laser must be calibrated to bring the rotating Z beam parallel to the
plumb line.
1) Turn off the laser. Press and hold the Auto/Man key on the laser.
2) Hold AUTO/MAN key. While holding, momentarily press the ON key.
Look for:
Y/Z LED flashes, then X LED flashes. Release Auto/Man key.
Y/Z LED will blink rapidly, indicating leveling. When the LED blinks
slowly, the laser is ready for Z axis calibration.
3) Use the arrow keys to align scanning beam with string plumb line.
4) Move the beam slightly so that the beam is over the plumb line for the
final check.
5) Press the small dot key on the remote to save calibration. The beam
will go off.