3.2.5 BATTERY CHARGE LEVEL INDICATORS
The column with the “exposure” symbol indicates the charge level of the
Batteries used for radiographic operations (X-ray exposures) and the column
with the “motor” symbol indicates the charge level of the batteries used for the
Mobile motion (motors).
When plugged into the mains (with the Line Circuit Breaker ON and
the Emergency Switch-Off deactivated), the Batteries automatically charge.
The color Indicators on both columns illuminate and scroll from the current
Generator battery charge level to 100%, until the Batteries are fully charged.
During the charging process both columns scroll up from the same level.
The Batteries require approximately 9 hours for a fully charge. To
charge the Batteries, there is not need to have the Console turned
ON. When the Batteries are fully charged, the Battery charge level
Indicators on both columns stop scrolling and only the Upper
Green Indicators remain illuminated.
When unplugged from mains, the Batteries discharge independently depending
on their use (X-ray exposures or motors) since the Mobile is provided with two
independent battery modules.
Upon disconnecting the Unit from the mains, if the Unit has been
connected for a short period of time, after several exposures or
after one heavy duty exposure, the Batteries need at least
30 seconds to stabilize the charge, after which the c orrect charge
level is shown on the Indicator.
The Battery Charge Level Indicators can be:
MOBILE UNIT
PLUGGED INTO MAINS
MOBILE UNIT
UNPLUGGED FROM MAINS
Key in “OFF”or“ON” position Key in “OFF” position
Key in “ON” position and
Console turned ON
Key in “ON” position and
Console turned OFF
Both Columns are scrolling as
described in the following Table.
Both Columns are OFF.
Each Column shows the respective
Battery charge level as described
in the following Table.
Only the Motors Column shows the
respective Battery charge level as
described in the following Table.
Note .