EasyManua.ls Logo

Agilent Technologies 3000 Series - Page 107

Agilent Technologies 3000 Series
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service
DC Gain Accuracy Test
4-11
13 Select the Vavg measurement as shown below.
14 Record the DMM voltage reading as V
DMM+
and the oscilloscope Vavg
reading as V
Scope+
in the DC Gain Test section of the Performance Test
Record.
15 Repeat step 14 for the remaining vertical sensitivities for channel 1 in
the DC Gain Test section of the Performance Test Record.
16 Set the power supply voltage to +6 mV.
17 Move the BNC cable on channel 1 to channel 2.
18 Press the
Save/Recall front panel button.
19 Select the
Storage item in the Save/Recall menu until Setups appears.
20 Select
Default Setup in the Save/Recall menu.
21 Set the channel 2 vertical sensitivity value to 2 mV/div.
22 Press the
Meas button on the front of the oscilloscope.
23 Select the
Voltage menu item.
24 Select the
Vavg measurement.
25 Record the DMM voltage reading as V
DMM-
and the oscilloscope Vavg
reading as V
Scope-
in the DC Gain Test section of the Performance Test
Record.
26 Repeat step 25 for the remaining vertical sensitivities for channel 2 in
the DC Gain section of the Performance Test Record.
Vavg
measurement

Table of Contents

Related product manuals