EasyManua.ls Logo

Agilent Technologies 3000 Series - Page 12

Agilent Technologies 3000 Series
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started
Compensating Probes
1-8
2 Set the square wave generator to a frequency of 1 MHz and an amplitude
of 1 Vp-p.
3 Press the
Autoscale front panel button.
Figure 1-4
Probe Compensation
4 If the waveform does not appear like the Correctly Compensated
waveform shown in Figure 1-4, then use a nonmetallic tool to adjust the
2 high frequency compensation adjustments on the probe for the flattest
square wave possible.
Over Compensated
Under Compensated
C
orrectly
C
ompensated
High frequency compensation adjustments

Table of Contents

Related product manuals