Operation 2
3200P pH Meter User Guide 29
To set up the system (System Setup)
System setup includes the following items:
1 Press [Setup].
2 Select System Setup, and press [Enter] to enter System Setup.
See Figure 14.
Calib Interval Calib Interval is the time interval between
recalibration prompts. The meter counts
the time since the last calibration and if
the time counted exceeds the entered
calibration time interval, the meter
displays a popup window to remind the
user to recalibrate the electrode. Calib
Interval is measured in hours. Input zero
(0) hours to disable.
Operator No Operator No is a three- digit number in the
range 000 to 200. Every set of recorded
data contains the Operator No. field. This
field is used to record the operator
performing measurements for regulated
(GLP) laboratories. If you need to track
the operator, assign a unique number for
each operator.
Time System time and date. See “To set the date
and time” on page 30.
Auto Delete Enable to automatically overwrite the
oldest stored data entries with new data if
the data storage is full.

Table of Contents

Other manuals for Agilent Technologies 3200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3200 Series and is the answer not in the manual?

Agilent Technologies 3200 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model3200 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals