139
Chapter 3 Features and Functions
Sum Modulation (Sum)
3
Sum Modulation (Sum)
In Sum Modulation (Sum), the modulating signal is added to the carrier
waveform. A typical use is adding gaussian noise to a carrier signal. The
waveform generator will accept an internal or external modulation
source.
For more information on the fundamentals of a sweep, refer to chapter 5,
“Tutorial.”
To Select Sum Modulation
The waveform generator allows a modulating signal to be added to
the carrier as a percentage of the carrier waveform amplitude in
resolution of 0.01%.
The waveform generator will allow Sum modulation to be added to
any carrier waveform.
Front-Panel Operation: You must enable Sum before setting up any
of the other modulation parameters. Press
,
then select Square.
Press
and then press
Type
and
Sum
.
At this point, you can
specify the other modulation parameters. Once you have done so,
press
Modulate
to turn on modulation.
Remote Interface Operation: To avoid multiple waveform changes,
enable Sum after you have set up the other modulation parameters.
SUM:STATe {OFF|ON}

Table of Contents

Other manuals for Agilent Technologies 33522A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33522A and is the answer not in the manual?

Agilent Technologies 33522A Specifications

General IconGeneral
BrandAgilent Technologies
Model33522A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals