106 34410A/11A/L4411A Service Guide
4 Disassembly and Repair
Repackaging for Shipment
If the unit is to be shipped to Agilent for service or repair, be sure to:
• Attach a tag to the unit identifying the owner and indicating the
required service or repair. Include the model number and full serial
number.
• Place the unit in its original container with appropriate packaging
material for shipping.
• Secure the container with strong tape or metal bands.
If the original shipping container is not available, place your unit in a
container which will ensure at least 4 inches of compressible packaging
material around all sides for the instrument. Use static–free packaging
materials to avoid additional damage to your unit.
Agilent suggests that you always insure shipments.
Cleaning
Clean the outside of the instrument with a soft, lint–free, slightly
dampened cloth. Do not use detergent. Disassembly is not required or
recommended for cleaning.
To Replace the 34410A/11A Power Line Fuse
The power line fuse is located within the multimeter’s fuse–holder
assembly on the rear panel (see page 50). The multimeter is shipped from the
factory with a power–line fuse installed. The supplied fuse is a 250mA
, 250V,
slo
w–blow, 5x20mm fuse, Agilent part number 2110–0817. If you determine
that the fuse is faulty, replace it with one of the same size and rating.