18 34410A/11A/L4411A Service Guide
1 Specifications
These specifications apply when using the 34410A/11A/L4411A multimeter
in an environment that is free of electromagnetic interference and
electrostatic charge.
When using the multimeter in an environment where electromagnetic
interference or significant electrostatic charge is present, measurement
accuracy may be reduced. Particularly note:
• The voltage measurement probes are not shielded and can act as
antennas, causing electromagnetic interference to be added to the signal
being measured.
• Electrostatic discharges of 4000 V or greater may cause the multimeter
to temporarily stop responding, resulting in a lost or erroneous reading.
Specifications are subject to change without notice. For the latest
specifications, see the product datasheet on the Web. Firmware updates
may also be available on the Web. Start at the appropriate product page:
www.agilent.com/find/34410A
www.agilent.com/find/34411A
www.agilent.com/find/L4411A
The specifications on the following pages are valid for Agilent 34410A, 34411A, or L4411A
multimeters with firmware revision 2.05, or later, installed.