34410A/11A/L4411A Service Guide 31
Specifications 1
Remote Interfaces GPIB IEEE–488, 10/100Mbit LAN, USB 2.0 Standard
Language SCPI – 1994.0, IEEE–488.2
LXI Compliance LXI Class C, Version 1.0
Warm–up Time 90 minutes
General Specifications (L4411A)
Power Supply: Universal 100V to 240V ± 10%
Power Line Frequency: 45 Hz to 440 Hz ± 10% automatically sensed
Automatically sensed at power–on, 400 Hz defaults to 50Hz.
Power Consumption: 50 VA peak (18 W average)
Operating Environment: Full accuracy for 0 °C to 55 °C
Full accuracy to 80% R.H. at 40
o
C Non–condensing
Storage Temperature –40 °C to 70 °C
Dimensions (HxWxL): 40.9 X 212.3 X 363.2mm
1.61 X 8.36 X 14.3 in
Weight: 1.9 kg (4.25 lbs)
Display: Dual-line, 5 X 7 dot matrix, 16 character LCD
Safety: Conforms to CSA, UL/IEC/EN 61010-1
EMC: Conforms to IEC/EN 61326-1, CISPR 11
Warrenty: 1 year
Triggering and Memory
Reading Hold Sensitivity: 1% of reading
Samples per Trigger: 1 to 50,000 (34410A)
1 to 1,000,000 (34411A/L4411A)
Trigger Delay: 0 to 3600 sec (20 μs step size)
External Trigger Low–power TTL compatible input programmable edge triggered
Delay: < 1us
Jitter: < 1us
Max rate up to 5 kHz
Min Pulsewidth 1 us
Voltmeter Complete 3 V Logic output
Polarity Programmable edge pulse
Pulsewidth Approximately 2 μs
Non–volatile Memory 50,000 reading
Sample Timer
Range Up to 3600 sec in 20 μs steps
Jitter < 100 ns