EasyManuals Logo

Agilent Technologies 4395A Service Manual

Agilent Technologies 4395A
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
2.
SOURCE LEVEL
ACCURA
CY/FLA
TNESS
TEST
Lev
el Accuracy
Minim
um
Limit
T
est
Result
Maxim
um
Limit
Measurement
Uncertain
t
y
0
1.0
dB
1.0
dB
6
0.12
dB
Lev
el
Flatness
F
requency
Minim
um
Limit
T
est
Result
Maximum
Limit
Measuremen
t
Uncertain
t
y
10
Hz
0
2.0
dB
2.0
dB
6
0.12
dB
50
Hz
0
2.0
dB
2.0
dB
6
0.12
dB
500
Hz
0
2.0
dB
2.0
dB
6
0.12
dB
2
kHz
0
2.0
dB
2.0
dB
6
0.12
dB
12 kHz
0
2.0
dB
2.0 dB
6
0.12
dB
60
kHz
0
2.0
dB
2.0
dB
6
0.12
dB
160
kHz
0
2.0 dB
2.0
dB
6
0.12 dB
500
kHz
0
2.0
dB
2.0
dB
6
0.13
dB
1
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
3
MHz
0
2.0 dB
2.0
dB
6
0.15 dB
10
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
45
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
75
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
110
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
140
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
190
MHz
0
2.0
dB
2.0
dB
6
0.15
dB
235
MHz
0
2.0 dB
2.0
dB
6
0.37 dB
290 MHz
0
2.0
dB
2.0 dB
6
0.37
dB
345 MHz
0
2.0 dB
2.0 dB
6
0.37 dB
400 MHz
0
2.0 dB
2.0 dB
6
0.37 dB
445 MHz
0
2.0 dB
2.0 dB
6
0.37 dB
470 MHz
0
2.0 dB
2.0 dB
6
0.37 dB
500 MHz
0
2.0 dB
2.0 dB
6
0.37 dB
4-2 Performance Test Record

Other manuals for Agilent Technologies 4395A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4395A and is the answer not in the manual?

Agilent Technologies 4395A Specifications

General IconGeneral
BrandAgilent Technologies
Model4395A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals