8.
INPUT CROSST
ALK TEST
10
Hz
Freq.
<
1 kHz
Measuremen
t
T
est
Result
T
est
Limit
R
in
to
A
Crosstalk
<
0
100
dB
R
in
to B
Crosstalk
<
0
100 dB
1 kHz
Freq.
<
100
kHz
Measuremen
t
T
est
Result
T
est
Limit
R
in
to
A
Crosstalk
<
0
100
dB
R
into
B Crosstalk
<
0
100
dB
100
kHz
Freq.
500
MHz
Measuremen
t
T
est
Result
T
est
Limit
Rin
to
A
Crosstalk
<
0
120
dB
R
in
to
B
Crosstalk
<
0
120
dB
9.
INPUT
IMPED
ANCE
TEST
100
kHz
Freq.
100 MHz
Measuremen
t
Test
Result T
est Limit
Measuremen
t
Uncertain
t
y
R
Return Loss
<
0
25
dB
6
3.16
dB
A
Return
Loss
<
0
25
dB
6
3.16 dB
B
Return
Loss
<
0
25
dB
6
3.16
dB
100
MHz
<
Freq.
500 MHz
Measuremen
t Test Result Test Limit Measuremen
t
Uncertaint
y
R Return Loss
<
0
15 dB
6
3.16 dB
A Return Loss
<
0
15 dB
6
3.16 dB
B Return Loss
<
0
15 dB
6
3.16 dB
Performance Test Record 4-7

Other manuals for Agilent Technologies 4395A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4395A and is the answer not in the manual?

Agilent Technologies 4395A Specifications

General IconGeneral
BrandAgilent Technologies
Model4395A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals