28 Site Preparation Guide
Site Preparation
Air Conditioning and Environmental Requirements
Air conditioning considerations include temperature, humidity, airborne dust,
and exhaust venting. Each of these is considered in more detail in the
following material.
Temperature, humidity, and altitude
The 5975 Series MSD is specified for operation under the following conditions:
Operation requires constant temperature
(variations <2 °C/h)
Operation and storage require a noncondensing, noncorrosive atmosphere
The temperature and humidity limits in Table 5
Maximum altitude for operation: 2300 m (7500 ft)
Environmental control systems must maintain these temperature and
humidity ranges.
The MSD is rated for 1200 Watts (4000 BTU/h). The GC and the data system
also contribute significantly to the cooling load although the exact amounts
depend on their configurations. Additional allowances should be made for
other heat sources such as heat from other equipment, heat from adjacent
rooms, and heat from laboratory personnel.
Table 5 Temperature and humidity limits
Temperature Humidity (relative)
Operation 15 °C to 35 °C (59 °F to 95 °F) 40% to 80%
Storage –20 °C to 70 °C (–4 °F to 158 °F) 0% to 95%
5975 Series Site Prep.book Page 28 Thursday, January 31, 2008 9:47 AM

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals