EasyManua.ls Logo

Agilent Technologies 6032A - Chapter 3 Manual Changes; Chapter 5 And 6 Manual Changes

Agilent Technologies 6032A
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
126
Chapter 3 Manual Changes:
Replace Figure 3-1 with the figure on the next page. SA Tables 3-5 through 3-7 do not apply to Option 001 units.
Chapter 5 and 6 Manual Changes:
On Page 94 under A3 Front Panel Board, the only part that applies to the Option 001 unit is R72, the OVP-adjust
potentiometer. Change R72 from 2100-1775 to 2100-4060.
On Page 102 under Chassis Electrical, add Neon Pilot Light, P/N 1450-0647. Also add OVP Cable (R72), P/N 06032-
60004. Under Chassis Mechanical, change screened front panel to:
06030-00010 (6030A)
06031-00011 (6031A)
06032-00021 (6032A)
On Figure 6-9, add the neon pilot light in series with R8, 33K, across pins 4 and 5 of transformer T3.
On Figure 6-12, the only part that applies to the Option 001 unit is R72, the OVP-adjust potentiometer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals