136
Performance Test Record - Agilent 6035A POWER SUPPLY (Page 2 of 2)
MODEL Agilent 6035A
Report No.______________ Date_____________________
Page
No.
Test Description
Minimum Spec. Results
*
Maximum
Spec.
Measurement
Uncertainty
Constant Voltage Tests
23 Voltage Programming
and Readback
Low Voltage (1.00V)
Front Panel Display
GPIB Readback
0.598
V
OUT
- 0.300
V
OUT
- 0.205
________V
________V
________V
1.400
V
OUT
+ 0.300
V
OUT
+ 0.205
12
µV
12
µV
12
µV
High Voltage (500V)
Front Panel Display
GPIB Readback
498.350
V
OUT
- 3.500
V
OUT
- 2.700
________V
________V
________V
501.650
V
OUT
+ 3.500
V
OUT
+ 2.700
6.5mV
6.5mV
6.5mV
23 Load Effect (200V)
V
OUT
- 0.060 _______mV V
OUT
+ 0.060 350µV
24 Source Effect (500V)
V
OUT
- 0.063 _______mV V
OUT
+ 0.063 800µV
25
26
PARD (Ripple and Noise)
RMS
Peak-to-Peak
0
0
_______mV
_______mV
0.050
0.160
7.5mV
4.8mV
27 Transient Recovery
(90% to 100% Full Load)
0 _______mV 0.200/5ms 24mV
Constant Current Tests
28 Current Programming
and Readback
Low Current (0.10A)
Front Panel Display
GPIB Readback
0.015
I
OUT
- 0.060
I
OUT
- 0.050
________A
________A
________A
0.185
I
OUT
+ 0.060
I
OUT
+ 0.050
56
µA
56
µA
56
µA
High Current (5.0A)
Front Panel Display
GPIB Readback
4.900
I
OUT
- 0.085
I
OUT
- 0.075
________A
________A
________A
5.100
I
OUT
+ 0.085
I
OUT
+ 0.075
2mA
2mA
2mA
29 Load Effect (2.0A)
I
OUT
- 0.034 ________mA I
OUT
+ 0.034 2µA
29 Source Effect (5.0A)
I
OUT
- 0.018 ________mA I
OUT
+ 0.018 2µA
30 PARD (Ripple and Noise)
0 ________mA 0.050 5mA