EasyManua.ls Logo

Agilent Technologies 6460 - Page 43

Agilent Technologies 6460
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Exercise 3 – Create a Triggered Dynamic MRM acquisition method
Task 1. Create a Triggered Dynamic MRM method from a Dynamic MRM method manually
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide 43
2 Change the method to a triggered
dynamic MRM method.
a Click the Acquisition tab in the QQQ
tab in the Method Editor.
b Mark the Triggered check box in the
Triggered MRM section. This section
is only available if the Scan Type is
Dynamic MRM.
c Select whether to automatically mark
the highest product ion as the Primary.
d Enter the value for Repeats.
Several columns are added to the
Scan segments table. These
columns only apply to a triggered
dynamic MRM method.
The value Repeats is the number of
times to acquire each of the
secondary transitions when the
triggering conditions are met.
3 Add additional transitions. See the
next image.
a Select the first compound. Right-click
and click Insert Row. Repeat to insert
another row
b Modify the information in these rows
to match the information in the next
image.
c Repeat for each compound.
For each compound, we are going
to add additional transitions.
4 Select the transitions that are the
Primary transitions. See the next
image.
a For each transition, mark the Primary
check box if it is a Primary transition.
b Verify that you have marked at least
one transition as the Primary
transition for each Compound Name.
You can select multiple transitions
from each compound to be Primary
transitions. If a transition has the
same Compound Name, then it is
part of the same compound. You
must mark at least one transition as
a Primary transition for each
compound.
Typically, the most abundant ion is
the Trigger transition.
Steps Detailed Instructions Comments

Related product manuals