Exercise 1 – Develop an acquisition method
Task 3. Find optimum fragmentor voltage for maximum response
14 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
Task 3. Find optimum fragmentor voltage for maximum response
Task 3 shows you how to carry out the optimization for fragmentor voltage by
creating selected ion-monitoring experiments for each compound within a
method and setting up multiple methods with varying fragmentor voltages.
Steps Detailed Instructions Comments
1 Set up six methods for six different
fragmentor voltages.
Change to a SIM experiment.
Use 60, 80, 100, 140, 180 and
220 volts as the fragmentor
voltages for the six methods.
Save the methods as
iiiMS2SIMxxx.m, where iii are
your initials and xxx is the
voltage.
a In the Scan Type dropdown list, click
MS2 SIM.