EasyManua.ls Logo

Agilent Technologies 7694 - Tightening the PTV Septumless Head Using the Leak Test Procedure

Agilent Technologies 7694
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintaining the Instrument
Tightening the PTV Septumless Head Using the Leak Test Procedure
113
Tightening the PTV Septumless Head Using the Leak
Test Procedure
If you are using a PTV inlet with a septumless head, you can use the leak test
procedure described in Carrier Gas Flow Path” on page 105 or “Carrier Gas
Flow Path” on page 110 to help you properly tighten the septumless head onto
the needle. As you tighten the fitting on the needle, monitor the flow rate on the
GC front display. When the flow rate indicates there is no leak, the connection
is tight enough.

Table of Contents

Other manuals for Agilent Technologies 7694

Related product manuals