EasyManuals Logo

Agilent Technologies 7694 User Manual

Agilent Technologies 7694
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
Troubleshooting
Vial Handling Problems (G1289/G1290 only)
119
Unwanted Peaks Present
Possible Causes
Dirty vials or septa.
Dirty laboratory air sealed in the vial with sample.
Carryover from the previous sample.
Contaminated GC system.
Dirty column and/or ferrules.
Suggested Actions
1. Bake the vials and septa in oven.
2. Flame the vials before use.
3. Wash the vials and septa and then dry them in the oven.
4. Purge the vial with argon during sample preparation.
5. Steam clean the headspace sampler. See “Steam Cleaning the Pneumatic
System” on page 95.
6. Perform a blank GC run; clean or bake out the GC.
7. Heat the graphite GC–column ferrules with a flame before use.
8. Call Agilent service.
Vial Handling Problems (G1289/G1290 only)
Gripper Arm Doesn’t Pick Up Sample Vial
Probable Causes
Arm movements out of adjustment.
Spring lock and spacer out of adjustment.
Carousel out of adjustment. See “Checking carousel adjustment” on
page 101.
10 ml vials used without adapters.

Table of Contents

Other manuals for Agilent Technologies 7694

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7694 and is the answer not in the manual?

Agilent Technologies 7694 Specifications

General IconGeneral
BrandAgilent Technologies
Model7694
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals