Using the Adv Fctn Key
Advanced Function Descriptions
52
or
VIAL EQ. INCR 0.0
Change the increments from the numeric keypad and then press the ENTER
key to execute the change in value.
G1289/G1290 Adv Fctn, 2, Enter
Select Parameter Incr. and use the cursor keys to change the vial equilibrium
time step (increment) and the oven temperature step:
VIAL EQUIL STEP 0.0
The cursor keys scroll to the next display:
OVEN TEMP STEP 0
PRESSURE UNITS
G1883 Adv Fctn, 5, Enter
G1289/G1290 Adv Fctn, 6, Enter
This function changes the pressure units for the displays among psi, kPa, and
bar. Pressing Adv Fctn, 6, and Enter displays the following message:
PRESSURE UNITS PSI
Psi is the default for pressure units when the sampler unit is powered up. Use
the cursor keys to scroll to kPa or bar. When units are shown on the display,
they are active.