Replacing the NPD Bead
Instruction Guide 11
6 Mount the new bead assembly on the NPD lid. Be careful not
to bump the bead on the sides of the lid or collector.
7 Replace the screws. Finger- tighten the first screw; tighten
the remaining screws normally and then completely tighten
the first screw. Do not overtighten the screws.
8 Carefully bend the bead assembly cable 90 °.