Measurement Functions – The SENSe Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 109
command:
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength?
syntax: :SENSe[n]:[CHANnel[m]]:POWer:WAVelength?[<wsp>MIN|MAX|DEF]
description: Inquires the current sensor wavelength.
parameters: none
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half
the sum of, the minimum programmable value and the
maximum programmable value
response: The wavelength as a float value in meters.
example sens1:pow:wav? +1.55000000E-006<END>
affects: All power meters
dual sensors: Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
description: For all sources, overwrites the current calibration values with the factory-set calibration set-
tings. See “:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination” on
page 110 and “:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]?” on
page 112 for information on calibrating your return loss module.
parameters: none
response: none
example sens1:ret:cal:fact
affects: All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
description: For the currently selected source, start the calibration and save the calibration values for a
defined reflectance reference measurement. See
“:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]” on page 112 for infor-
mation on setting the return loss value of your reference reflector.
parameters: none
response: none
example sens1:ret:cal:coll:refl
affects: All return loss modules

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals