How to Co-ordinate Two Modules VISA Programming Examples
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 199
How to Co-ordinate Two
Modules
This example shows the interaction of two modules in the same frame.
Install a Power Sensor in Slot 1 and a Laser Source in Slot 2 and connect
the Laser Source output to the Power Sensor input, before executing this
example.
#include <stdio.h>
#include <stdlib.h>
#include <string.h>
#include <visa.h>
/* function prototypes for this examples */
/* function for a simple error handling explained in example 1 */
void checkError(ViSession session, ViStatus err_status );
void main (void)
{
ViStatus errStatus; /* returned error code from visa call */
ViSession defaultRM; /* default visa resource manager variable
*/
ViSession vi; /* current session handle */
ViChar replyBuf[256]; /* buffer holding answers of the
instrument */
ViChar c; /* used in the keyboard wait loop */
ViInt32 i; /* loop counter */
ViInt32 cmdDone; /* return value for OPC command */
/* First get initialized the visa library (see example 1) */
errStatus = viOpenDefaultRM (&defaultRM);
if(errStatus < VI_SUCCESS)
{
printf("Failed to open VISA Resource manager\n");
exit(errStatus);
}
/* Open session to GPIB device at address 20; */

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals