EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8163B

Agilent Technologies 8163B Programming Guide

Agilent Technologies 8163B
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
The Status Model Introduction to Programming
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 35
Status System for 8163A/B & 8164A/B
The status system for the Agilent 8163A/B Lightwave Multimeter and the
Agilent 8164A/B Lightwave Measurement System returns the status of 2
and 5 module slots respectively. The Operational/Questionable Status
Summary Registers consist of one level and are described by Figure 4 .
Any commands that require LEVel1 do not apply to these mainframes.
Figure 4 The Operational/Questionable Status System for 8163A/B & 8164A/B
Operational/Questionable Status
Operational/Questionable Status
Operational/Questionable Status
Condition Summary Register
Event Summary Register
Enable Summary Mask
Register
Status Byte
Status Byte Register
Status Summary
to next
Operational/Questionable
Operational/Questionable
Operational/Questionable
Slot Status Condition
Slot Status Event
Slot Status Enable Mask
&
&
&
&
OR
level
to next
for a positive
&
&
&
&
OR
level
Slot 1
transition
for a positive
transition
Register
Register
Register

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals