Measurement Operations & Settings Root Layer Command
80 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Root Layer Command
The commands in the Slot subsystem allow you to query the following:
a particular slot, for example, using slot1:empt?,
or, an Optical Head attached to an Optical Head Interface Module, for
example, an Optical Head Interface Module in slot1 with an Optical
Head attached to channel 2, using slot1:head2:empt?.
command:
:LOCK
syntax: :LOCK<wsp><boolean>, <value>
description: Switches the lock off and on.
High power lasers cannot be switched on, if you switch the lock on. High power lasers are
switched off immediately when you switch the lock on.
parameters: A boolean value: 0 or OFF: switch lock off
1 or ON: switch lock on
<value> is the four-figure lock password.
response: none
example: lock 1,1234 - 1234 is the default password
command:
:LOCK?
syntax: :LOCK?
description: Queries the current state of the lock.
parameters: none
response: A boolean value: 0: lock is switched off
1: lock is switched on
example: lock? 1<END>

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals