Measurement Operations & Settings Signal Conditioning
156 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:OUTPutn[:CHANnel[m]]:POWer:UNit?
syntax: :OUTPutn[:CHANnel[m]]:POWer:UNit?
description: Queries whether the power unit is dBm or W
parameters: none
response: boolean 0 The power unit is dBm
1 The power unit is W.
example: OUTP1:POW:UN? 0<END>
affects: All attenuator modules.
command:
:OUTPutn[:CHANnel[m]]:[STATe]
syntax: :OUTPutn[:CHANnel[m]]:[STATe]<wsp>OFF(0) | ON(1)
description: Sets the state of the shutter.
parameters: OFF or 0
ON or 1
Shutter closed.
Shutter open.
response: none
example: OUTP1:STAT OFF
affects: All attenuator modules.
command:
:OUTPutn[:CHANnel[m]]:[STATe]?
syntax: :OUTPutn[:CHANnel[m]]:[STATe]?
description: Queries the state of the shutter.
parameters: none
response: boolean 0 The shutter is open.
1 The shutter is closed.
example: OUTP1:STAT? 0<END>
affects: All attenuator modules.
command:
:OUTPutn[:CHANnel[m]]:STATe:APOWeron
syntax: :OUTPutn[:CHANnel[m]]:STATe:APOWeron<wsp>OFF(0) | ON(1)
description: Sets the state of the shutter when the mainframe is turned on.
parameters: OFF or 0
ON or 1
Shutter closed after mainframe power on.
Shutter open after mainframe power on.
response: none
example: OUTP1:APOW OFF
affects: All attenuator modules.

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals