Measurement Operations & Settings Root Layer Command
84 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:CSV?
syntax: :SLOT[n]:HEAD[m]:WAVelength:RESPonse:CSV?
description: Returns the wavelength response from the attenuator module in CSV format.
response: Wavelength Response table as a string
response
format:
The string is a comma separated value (CSV) list and can be written to a file and be pro-
cessed with a spreadsheet program.
List format:
λ1, c1\n
λ2, c2\n
.......
λn, cn\n
"," separates wavelength and response factor
"\n" = ASCII code 10 separate value pairs
example: slot1:head1:wav:resp:csv? 1200e-6,2.019\n 1210e-6,
1.956\n...
affects: Attenuator with power control, all powermeters, return loss modules
command:
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:SIZE?
syntax: :SLOT[n]:HEAD[m]:WAVelength:RESPonse:SIZE?
description: Returns the number of elements in the wavelength response table.
response Number of elements in the wavelength table as an integer value
example: slot2:head1:wav:resp:size? 50<END>
affects: Attenuator with power control, all powermeters, return loss modules
command:
:SPECial:REBoot
syntax: :SPECial:REBoot
description: Reboots the mainframe and all modules.
parameters: none
response: none
example: spec:reb

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals