130 Chapter2
Making Adjustments
19. Entering External ALC Correction Constants for Option 010 or 011
4. Press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4, SERVICE CAL
EXECUTE TITLE CAL, More 1 of 4, More 2 of 4,
SERVICE DIAG, DISPLAY CAL DATA, NEXT PAGE
Verify that the XSLOP and XOFST values are the same as you
entered in step 3.
5. Perform the “Checking the External ALC for Option 010 or 011"
procedure.