1-80
Making Measurements
Using Limit Lines to Test a Device
1. To offset all of the segments in the limit table by a fixed frequency, (for example,
3 MHz), press:
The analyzer beeps and a FAIL notation appears on the analyzer display, as shown in
Figure 1-63.
Figure 1-63 Example Stimulus Offset of Limit Lines
To return to 0 Hz offset, press:
To offset all of the segments in the limit table by a fixed amplitude, press:
The analyzer beeps and a FAIL notation appears on the analyzer display.
To return to 0 dB offset, press:
To offset the amplitude offset value by the active marker reading, press
. Pressing shows the current value.
System
LIMIT MENU
LIMIT LINE
LIMIT LINE OFFSETS
STIMULUS OFFSET
3 M/µ
STIMULUS OFFSET
0 x1
AMPLITUDE OFFSET
5 x1
AMPLITUDE OFFSET
0 x1
MARKER
AMP. OFS.
AMPLITUDE OFFSET