EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8960 Series

Agilent Technologies 8960 Series Programming Guide

Agilent Technologies 8960 Series
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
45
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Proguide\Chapters\prog_prog_guide_step6.fm
Step 6: Make Measurements
Example 1. Programming Example:
The following program segment illustrates making a transmit power measurement and a phase and frequency
error measurement concurrently using the recommended process shown in “Step 6: Figure 1. Process for
Making Measurements” on page 44.
!**********************************************************************
! Step 6: Make Measurements
!**********************************************************************
!
! Step 6a: Start Set of Concurrent Measurements:
!
OUTPUT Test_set;”INIT:TXP;PFER”
!
! Step 6b: Determine If A Measurement Is Done:
!
LOOP
OUTPUT Test_set;”INIT:DONE?”
ENTER Test_set;Meas_done$
!
! Step 6c: Obtain Measurement Results
!
SELECT Meas_done$
CASE “TXP”
OUTPUT Test_set;”FETC:TXP:POW?”
ENTER Test_set;Avg_tx_power
CASE “PFER”
OUTPUT Test_set;”FETC:PFER:RMS?”
ENTER Test_set;Max_rms_phas_er
END SELECT
EXIT IF Meas_done$ = “NONE”
END LOOP

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals