Chapter 3 Calibration Procedures
Automating Calibration Procedures
46
Automating Calibration Procedures
You can automate the complete verification procedures outlined in this chapter 
if you have access to programmable test equipment. You can program the 
instrument configurations specified for each test over the remote interface. 
You can then enter readback verification data into a test program and compare 
the results to the appropriate test limit values.
You can also enter calibration constants from the remote interface. Remote 
operation is similar to the local front-panel procedure. You can use a computer 
to perform the adjustment by first selecting the required setup. The calibration 
value is sent to the power supply and then the calibration is initiated over the 
remote interface. The power supply must be unsecured prior to initiating the 
calibration procedure. An example program of Excel 97 for calibration over 
the GPIB interface is listed at the end of this chapter.
For further details on programming the power supply, see chapters 3 and 4 in 
the Agilent E3633A and E3634A User’s Guide.
Test Considerations
To ensure proper instrument operation, verify that you have selected the 
correct power-line voltage prior to attempting any test procedure in this 
chapter. See page 24 in chapter 2 for more information.
Ensure that all connections of terminals (both front panel and rear panel) are 
removed while the power supply internal self-test is being performed.
For optimum performance verification, all test procedures should comply with 
the following recommendations:
• Assure that the calibration ambient temperature is stable and between 20°C 
and 30°C.
• Assure ambient relative humidity is less than 80%.
• Allow a 1-hour warm-up period before verification or calibration.
• Use short cables to connect test set-ups.
Caution   The tests should be performed by qualified personnel. During performance 
verification tests, hazardous voltages may be present at the outputs of the power 
supply.