*TST?
This query performs a
complete
self-test of the power supply. Returns0’’ if
the self-test passes or ‘1’’ or any non-zero value if it fails. If the self-test fails,
an error message is also generated with additional information on why the test
failed.
*SAV { 1 | 2 | 3 }
This command stores the present state of the power supply to the specified
location in
non-volatile
memory. Three memory locations (numbered 1, 2 and
3) are available to store operating states of the power supply. The state storage
feature ‘‘remembers’’ the states or values of the following commands:
CURR, CURR:STEP, CURR:TRIG, CURR:PROT, CURR:PROT:STAT
DISP, OUTP, OUTP:REL, TRIG:DEL, TRIG:SOUR, VOLT,
VOLT:STEP, VOLT:TRIG, VOLT:PROT, VOLT:PROT:STAT,
and
VOLT:RANG
To recall a stored state, you must use the same memory location used
previously to store the state.
*RCL { 1 | 2 | 3 }
This com mand recall s a previousl y stored state. To recall a stored state, you
must use the same mem ory location used previously to store the state.
Note DISP {OFF|ON} can be stored and recalled in remote interface mode only. Going
to local mode automatically sets the display state to ON.

Table of Contents

Other manuals for Agilent Technologies E3634A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3634A and is the answer not in the manual?

Agilent Technologies E3634A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3634A
CategoryPower Supply
LanguageEnglish

Related product manuals