Chapter 5 Error Messages
Calibration Errors
134
Calibration Errors
The following errors indicate failures that may occur during a calibration. Refer to
the Service Information for more information.
701 Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the power
supply. When applicable, this error will occur at power-on to warn you that the power
supply is unsecured.
702 Cal secured
The power supply is secured against calibration.
703 Invalid secure code
An invalid calibration security code was received when attempting to unsecure or
secure the power supply. You must use the same security code to unsecure the power
supply as was used to secure it, and vice versa. The security code may contain up to
11 alphanumeric characters.
704 Secure code too long
A security code was received which contained more than 12 characters.
705 Cal aborted
A calibration in progress is aborted when you press any front-panel key, send a device
clear, or change the local/remote state of the instrument.
706 Cal value out of range
The specified calibration value (CALibration:VALue) is not valid for the present
measurement function and range.
708 Cal output disabled
Calibration is aborted by sending OUTP OFF command during calibrating a output.
712 Bad DAC cal data
The specified DAC calibration values (CAL:VOLT or CAL:CURR) are out of range.
Note that the new calibration constants are not stored in the non-volatile memory.