Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
57
3
4 Clear the overvoltage condition and exit this menu.
The annunciator for OVP operation will not flash any more. The output will return to
meter mode.
1 Raise the OVP trip level higher than the level tripped.
2 Select the OVP CLEAR mode by turning the knob.
3 Clear the overvoltage condition and exit this menu.
The annunciator for OVP operation will not flash any more. The output will return to
the meter mode.
• Remote interface operation:
VOLT:PROT {<voltage>|MIN|MAX} Set the OVP level
VOLT:PROT:STAT {OFF|ON) Disable or enable the OVP circuit
VOLT:PROT:CLE Clear the tripped OVP circuit
done
By Adjusting OVP trip level
OVP ON
OVP CLEAR
done