EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4445A

Agilent Technologies E4445A Manual

Agilent Technologies E4445A
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
134
Front-Panel Key and SCPI Command
Reference
4.2.2.3.10 Serial No.
Returns the serial number of the power meter.
4.2.2.3.11 Show Setup
Display the current configuration information for the power meter.
4.2.2.3.12 Time Out
Sets and Gets the time-out time for connecting the power meter in units of seconds and milliseconds.
Mode: MRECEIVE
Remote Command: :SYSTem:COMMunicate:PMETer:SNUMber?
Preset: [NONE]
State Saved: Saved in instrument state.
Remote Command Notes: Global to the current mode.
Example: :SYST:COMM:PMET:SNUM?
Mode: MRECEIVE
Dependencies/Couplings: Enabled only when the power meter is available.
Key Path:
System, More, More, Power Meter, Power Meter Config
Mode: MRECEIVE
Key Path:
System, More, More, Power Meter, Power Meter Config, More
Remote Command: :SYSTem:COMMunicate:PMETer:TCONnect:TOUT <time>
:SYSTem:COMMunicate:PMETer:TCONnect:TOUT?
Unit: s | ms
Range: 1.0s to 120.0s
Preset: 10.0s
State Saved: Saved in instrument state.
Remote Command Notes: Global to the current mode.
Example: :SYST:COMM:PMET:TCON:TOUT 18
:SYST:COMM:PMET:TCON:TOUT?

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals