Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
144
Front-Panel Key and SCPI Command
Reference
4.2.5 Input
This key provides the same behavior as that of Input key of the other options except the options for Input Port.
Under Mode Setup menu, parameters can be changed via using the Form.
4.2.5.1 Input Port
This key provides the same behavior as that of Input Port key of the other options except the new options for
“Audio” and “Audio Calibrator”.
4.2.5.2 Ext Atten
Sets the loss equal to the external attenuation used when measuring the device under test.
Mode: MRECEIVE
Key Path:
Input, Input Port
Remote Command: [:SENSe]:FEED RF|AREFerence|AUDio|AFALign
[:SENSe]:FEED?
Preset: RF
State Saved: Saved in instrument state.
Range: RF | Amptd Ref | Audio | AF Calibrator
Dependencies/Couplings: The AF and AF Calibrator keys are blank when opt. 107 is not installed.
Remote Command Notes: Global to the current mode.
Example: :FEED Audio
:FEED?
Mode: MRECEIVE
Key Path:
Input, Ext Atten
Remote Command: [:SENSe]:CORRection [:RF]:LOSS <rel_power>
[:SENSe]:CORRection [:RF]:LOSS?
Preset: 0 dB
State Saved: Saved in instrument state.
Range: − 50 to +50 dB
Remote Command Notes: You must be in the Basic mode to use this command. Use INSTrument:SELect
to set the mode. Value is global to Basic mode.
Example: POW:RF:GAIN ON
POW:RF:GAIN?