4.2.2.3.4 Model No.
Returns the model number of the power meter.
4.2.2.3.5 Power Meter GPIB Address
Sets the GPIB address of the power meter.
4.2.2.3.6 Power Meter Firmware Version
Returns the firmware version of the power meter connected.
Mode: MRECEIVE
Remote Command: :SYSTem:COMMunicate:PMETer:MODel?
Preset: [NONE]
State Saved: Saved in instrument state.
Remote Command Notes: Global to the current mode.
Example: :SYST:COMM:PMET:MOD?
Mode: MRECEIVE
Key Path:
System, More, More, Power Meter, Power Meter Config
Remote Command: :SYSTem:COMMunicate:PMETer:GPIB:ADDRess <integer>
:SYSTem:COMMunicate:PMETer:GPIB:ADDRess?
Range: 1 to 30
Preset: 13
State Saved: Saved in instrument state.
Remote Command Notes: Global to the current mode.
Example: :SYST:COMM:PMET:GPIB:ADDR 18
:SYST:COMM:PMET:GPIB:ADDR?
Mode: MRECEIVE
Remote Command: :SYSTem:COMMunicate:PMETer:FW:VERSion?
Preset: [NONE]
State Saved: No
Remote Command Notes: Query Only
Example: SYST:COMM:PMET:FW:VERS?

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals